YB138R01K-重金属镉_组织_0.1ppm_卡壳_KJ50918.docx
Heavy metal cadmiumInstructions for use of the quick test card
Product number:YB138R01K
1 Principle and use
This product uses the immune competition method to quickly detect the content of heavy metal cadmium in aquatic products (fish, shrimp, crab, shellfish), which is suitable for on-site rapid testing of various enterprises, testing institutions and supervisory departments.
2 Technical indicators ExamineCard detection limit:Fish meat 0.1 Medium sizeG/kg (PpMedium size), crustaceans (excluding sea crabs and shrimps) 0.5 Medium sizeG/kg (PpMedium size), double-shelled shellfish2 Medium sizeG/kg (PpMedium size), sea crab, shrimp aunt 3 Medium sizeG/kg (PpMedium size)
3 You need to bring your own equipment and reagents
(1)Instruments: homogenizer, nitrogen drying device, oscillator, centrifuge, scale pipette, balance (sensing 0.01g)
(2)Micropitte: single channel 20 µL-200 µL, 100µL-1000µL
Four Sample pre-processing
Note: The sample needs to be shelled, peeled, internal organs, shrimp head, crab yolk, crab paste, homogeneous sample
Fish meat : Weigh 1 G homogeneous sample (minced meat or meat puree) to 5 Milliliters In the centrifuge tube, add 2 mL reagent A, add 0.2 mL reagent B, violent vortex oscillation into a paste, centrifuge at 4000rpm at room temperature for 3 minutes; take the centrifuged supernatant 200μL to 1.5 Milliliters In the centrifuge tube, add 100μL reagent C and mix well. This mixture is the liquid to be tested.
Crustaceans (excluding sea crabs and shrimps) : Weigh 1 G homogeneous sample (minced meat or meat puree) to 5 Milliliters In the centrifuge tube, add 2 mL reagent A, violent vortex oscillation into a paste, centrifuge at 4000rpm at room temperature for 3 minutes; take 200μL of the centrifugal supernatant to 1.5 Milliliters In the centrifuge tube, add 0.1 mL reagent B, mix up and down, wait for the reaction for 3 minutes, add 600μL reagent D and mix well. This mixture is the solution to be tested.
Double-shelled shellfish (remove internal organs) : Weigh 1 G homogeneous sample (minced meat or meat puree) to 5 Milliliters In the centrifuge tube, add 2 mL reagent A, add 0.2 mL reagent B, violent vortex oscillation into a slurry, centrifuge at 4000rpm at room temperature for 3 minutes; take the centrifuged supernatant 50μL to 1.5 Milliliters In the centrifuge tube, add 1 mL reagent D, mix well, this mixture is the liquid to be tested.
Sea crab, shrimp: Weigh 1g of homogeneous sample (minced meat or meat puree) to 5 Milliliters In the centrifuge tube, add 2 mL reagent A, violent vortex oscillation into a paste, centrifuge at 4000rpm at room temperature for 3 minutes; take the centrifugation supernatant 30μL to 1.5 Milliliters In the centrifuge tube, add 0.1 mL reagent B, mix up and down, wait for the reaction for 3 minutes, add 600μL reagent D and mix well. This mixture is the solution to be tested.
6 Result judgment

Negative(-):T line (test line, near the end of the sample hole) color development DeepAt PerhapsBe equal toThe color of the C line (control line) indicates that the concentration of the object to be tested in the sample is below the detection limit or there is no residue of the object to be tested.
Positive(Plus):The color rendering of the T line is lower than CentigradeLine color development, or TIf the line is not colored, it means that the concentration of the test in the sample is higher than the detection limit; the lighter the T line than the C line, the higher the concentration of the test in the sample.
Be invalid:Contrast line (C)Not yet A purple stripe appears, indicating that the operation process is incorrect or the detection card has been invalidated. In this case, it should be again Careful Read the instruction manual and use the new test card.Repeat New test.If the problem persists, do it immediately.Stop Stop using this batch number of products and contact the supplier..
7 Notes
(1) Please follow the operation steps to test. Do not touch the color display area of the test strip during operation, and avoid direct sunlight when testing.Blow directly with the electric fan.
(2) This test paper is a disposable product. Please do not reuse it.Or mix inspection cards from different batches.
(3)When there is a positive result, it is recommended to retest with this card.The test results of this product are for reference only. If you need to confirm, please refer to the relevant national standards and methods.
(4) If you need to directly test the standard product, you need to use the special diluent in the kit to make it.
(5) Please use it as soon as possible after the sample is processed, and the time process needs to be reprocessed before testing.
(6) Self-water, distilled water, pure water or deionized water cannot be used as negative controls.
8, storage conditions and validity period
Storage conditions: 4-30℃ avoid light Dry Keep it, don't freeze it.
Validity period:12 months.