Prochloraz Colloidal Gold Rapid Detection Card Instruction Manual
Product Number: YB160C01K
Summary
Prochloraz & Prochloraz-manganese chloride complex is an imidazole broad-spectrum fungicide. It has obvious control effect on diseases caused by ascomycetes and demagiacs in various crops. It can be used for the prevention and control of diseases such as food, vegetables, and fruit trees. It has no internal absorption effect and has certain conduction properties. Prochloraz and Prochloraz are toxic fungicides. They are toxic for swallowing and inhalation. They can penetrate through the skin and are extremely toxic to aquatic organisms. They may have long-term adverse effects on the water environment.
Detection principle
This product adopts the principle of competitive inhibition immunochromatography. Prochloramine in the sample binds to the specific antibody labeled with colloidal gold, which inhibits the binding of the antibody to the antigen on the NC film detection line (T line), resulting in the change of the color depth of the T line; according to the color development depth judgment result of the T line and the quality control line (C line).
Scope of application
This product is suitable for the qualitative detection of prochloramine residues in fresh vegetables, fruits and other samples.
Note: The type of test sample refers to the national standard GB2763.
Testing Limit
0 mg/kg (ppm)
Kit Composition
Serial Number
Specification
Composition
10 times/box
20 times/box
(1)
Test Card (1 T/pack)
10 Card
2 0 Card
(2)
Gold Standard Micropore (10 Bore/Cylinder)
10 Bore/Cylinder
20 Bore/Cylinder
(3)
Dilution (50 mL/bottle)
1 Bottle
2 Bottle
(4)
Disposable Straw
10 Piece
20 Piece
(5)
20 mL Sample Cup
1 Piece
2 Piece
Cleaning to avoid cross contamination.
(2) The test environment temperature should be controlled at 18~ 30 ℃, the temperature is too high or too low, will affect the test results.
(3) Please follow the test steps for testing. Do not touch the color-developing area of the test strip during operation to avoid direct sunlight and direct blowing of electric fans.
(4) Please use the sample as soon as possible after processing. If the time is too long, you need to re-process the sample for re-testing.
(5) The sample solution to be tested needs to be clarified, otherwise it will easily lead to abnormal phenomena such as insignificant color development, which will affect the judgment of experimental results.
(6) Products that have expired or damaged aluminum foil bags cannot be used. Please use the test card immediately after unpacking.
(7) This product is a one-time product, please do not reuse or mix test cards from different batches.
(8) When a positive result appears, it is recommended to re-test. The test results of this product are for reference only. If you need confirmation, please refer to the relevant national standards and methods.
(9) During product evaluation, if you need to directly test the standard product, you need to prepare the special dilution solution in the kit.
(10) Tap water, distilled water, purified water or deionized water cannot be used as a negative control.
Safety instructions
(1) The experiment needs to match the corresponding experimental equipment and wear the necessary experimental equipment (white clothes, gloves, masks, etc.).
(2) The test kit needs to be stored. Please put it in a place that is not easy for children to contact, and do not eat the reagents provided.
(3) All the items used in the experiment should be properly disposed of after application, and do not let children contact it.
(4) Keep the laboratory clean and tidy and the air circulation of the experimental environment after the experiment.
(5) The experimental waste should be collected separately, and it is recommended to dispose of it according to the medical waste.
(6) Do not eat the reagents provided.
storage conditions and valid period
(1) Storage conditions: 2-30 ℃ protected from light, do not freeze.
(2) valid period: 12 months.