Summary



Cyfluthrin (Pyrethroids) is a highly effective insecticide cyfluthrin insecticide, acaricide, with contact and repellent effect, can control vegetables, cotton, cereal crops Lepidoptera, Hemiptera and mite pests, widely used in various fruit trees, cotton, vegetables, tea and other crops insect mite control.


detection principle



This product adopts the principle of competitive inhibition immunochromatography. The cyfluthrin in the sample binds to the specific antibody labeled with colloidal gold, which inhibits the binding of the antibody to the antigen on the NC membrane detection line (T-line), resulting in the change of the color depth of the T-line; according to the color development depth judgment results of the T-line and the C-line.


Uses



used for the initial screening detection of cyfluthrin content in fresh vegetables, fruits and other samples.


The minimum detection limit of the sample



cyfluthrin: 0.2 mg/kg (ppm)


cross-reactivity and product performance



detection Added 3mg/kg of cyfluthrin, deltamethrin and other residual samples, the results were positive.


kit composition



serial number

specification
composition

1 0 times/box

20 times/box



(1)

test card (containing dropper, desiccant)

10 card

20 card



(2)

gold standard micropores (in aluminum foil bag)

10 hole

20 hole



(3)

dilution (60 mL/bottle)

1 bottle

2 bottle



(4)

20 mL sample cup with lid

1



(5)

1 5 mL centrifuge tube

10

20



(6)

manual

1

1 copies Precautions




(2) Sample requirements: Avoid spoilage and deterioration of the sample; Avoid large pieces of soil (can be thrown off or wiped with other clean items).


(3) Before testing, it is recommended that the sample be fully stirred and mixed (if the sample is less, the representative part should be taken, and then the sample should be weighed), so that the test results can more accurately reflect the actual drug residue of the sample.


(4) The detection environment temperature should be controlled at 20~ 30 ° C. Too high or too low temperature will affect the test results.


(5) Please follow the detection steps for testing. Do not touch the color-developing area of the test strip during operation, and avoid direct sunlight and electric fan direct blowing.


(6) The sample solution to be tested needs to be clarified, otherwise it will easily lead to abnormal phenomena such as insignificant color development, which will affect the judgment of the experimental results.


(7) expired or aluminum foil bag damaged products can not be used, please use the test card immediately after unpacking.


(8) Please use the sample as soon as possible after processing, if the time is too long, you need to reprocess the sample for re-testing.


(9) When positive results appear, re-testing is recommended. The test results of this product are for reference only. If you need confirmation, please refer to the relevant national standards and methods.


(10) When adding standard verification, the standard solution solvent generally chooses methanol, and the final sample is recommended to add scalar amount at 10-50 μL.


(11) When directly testing the standard, use the solvent kit supporting the extract, and the amount of organic solvent added is controlled within 1%. Tap water, distilled water, purified water or deionized water cannot be used as a negative control.


safety instructions



(1) The experiment needs to match the corresponding experimental equipment and wear the necessary experimental equipment (white clothes, gloves, masks, etc.).


(2) The test kit should be properly stored and placed in a place that is not easy for children to contact.


(3) Keep the laboratory clean and the air circulation of the experimental environment after the experiment.


(4) This product is a disposable product, and it should be properly disposed of after the test. The experimental waste is collected separately. It is recommended to be disposed of according to the medical waste.


(5) The reagents involved in this product are safe and reliable, and do not contain carcinogenic, highly toxic, flammable, explosive, and highly corrosive reagents, but they should not be eaten.


storage conditions and valid period



(1) Original packaging: dry storage at 2-30 ° C away from light, do not freeze, valid period 12 months.


(2) After unpacking: use the test card immediately after unpacking, do not freeze.