Albendazole Colloidal Gold Rapid Detection Card Instruction Manual
|
Product number: YB018R01K
|
summary
| |
Albendazole (Albendazole, ALB), also known as albendazole or albendazole, belongs to benzimidazole derivatives. It is a broad-spectrum deworming drug with high efficiency and low toxicity. It is currently widely used in the prevention and treatment of parasitic diseases in animals. Albendazole is a more toxic type of benzimidazole drugs. Although the application of therapeutic doses will not cause poisoning reactions, continuous overdose administration may sometimes cause serious adverse reactions, and continuous long-term use may also lead to cross-drug resistance. This product is suitable for rapid on-site detection in enterprises, testing institutions, supervision departments and other fields.
|
Detection principle
| |
This product applies the principle of competitive inhibition colloidal gold immunochromatography made , for detection organization In the sample Albendazole residue After the sample solution is dropped into the sample hole of the test card, the Albendazole It binds to the gold-labeled antibody, thereby preventing the gold-labeled antibody from interacting with the cellulose membrane Albendazole The conjugate binds. According to T line and C line color depth judgment results.
|
Scope of application
| |
This product is suitable for aquatic products, livestock and poultry tissue samples Albendazole Qualitative detection of residues.
|
Note: The types of test samples refer to the national standard GB2763-2021 。
|
detection limit
| |
Albendazole detection limit: 50 μg / kg (ppb)
|
Kit composition
| |
serial number
| Specifications
composition
| 10 times/box
| 20 times/box
| | | | |
(1)
| Detection card (Contains gold standard micropores, droppers, desiccant)
| 10 servings
| 20 servings
| | | | |
(2)
| Albendazole diluent
| 1 bottle
| 2 bottles
| | | | |
(3)
| 15mL centrifuge tube
| 2 sticks
| 4 sticks
| | | | |
( 4)
| manual
| 1 serving
| 1 serving
| | | | |
Precautions
| |
( 1) Before each sample inspection, the knives used to cut the sample should be cleaned to avoid cross-contamination.
|
( 2) The test environment temperature should be controlled at 18 to 30 degrees Celsius. If the temperature is too high or too low, it will affect the test results.
|
( 3) Please follow the detection steps for testing. Do not touch the color rendering area of the test strip during operation, and avoid direct sunlight and direct blowing of the electric fan.
|
( 4) Please use the sample as soon as possible after processing. If it takes too long, you need to reprocess the sample for testing.
|
( 5) The sample solution to be tested needs to be clarified, otherwise it will easily lead to abnormal phenomena such as insignificant color development, which will affect the judgment of the experimental results.
|
( 6) Products that have expired or the aluminum foil bag is damaged cannot be used. Please use the test card immediately after unpacking.
|
( 7) This product is a one-time product, please do not reuse or mix test cards from different batches.
|
( 8) When a positive result appears, it is recommended to retest. The test results of this product are for reference only. If you need confirmation, please refer to the relevant national standards and methods.
|
( 9) During product evaluation, if you need to directly test the standard product, you need to prepare it with a special diluent in the kit.
|
( 10) Tap water, distilled water, purified water, or deionized water cannot be used as a negative control.
|
safety instructions
| |
( 1) The experiment needs to match the corresponding experimental equipment and wear the necessary experimental equipment (white clothes, gloves, masks, etc.).
|
( 2) The test kit should be stored in a place that is not easily accessible to children.
|
( 3) All items used in the experiment should be properly disposed of after application.
|
( After the experiment, maintain the cleanliness of the laboratory and the flow of air in the experimental environment.
|
( 5) Experimental waste is collected separately, and it is recommended to treat it as medical waste.
|
( 6) Do not consume the reagents provided.
|
Storage conditions and valid period
| |
( Storage conditions: Store at 2-30 ° C away from light, do not freeze.
|
( 2) valid period: 12 months.
|